For example, using an external frequency of 1 Hz to open the gate and
selecting a time base of 10-6 to be passed to the DCUs, would result in a
count of 1,000,000  1 count.  The accuracy of the reading is now increased
to 0.0001 percent.  Compared to the frequency measurement of a 1 Hz signal 
with a possible 100 percent error, this is a great improvement.  The period
and frequency measurements are the two most widely used functions of an
electronic counter.
c. Totalizing.  An electronic counter can be operated in the totalizing
mode, illustrated in figure 5, with the main gate flip-flop controlled by 
the start-stop switch.  With the switch at START, the gate is opened and the
DCUs  will  totalize the  input  pulses until the  main  gate  is closed  by
changing the switch to STOP.  The display on the counter will then show the
pulses received during the interval between manual START and STOP.
Figure 5.
Totalize measurement.
d. Ratio Measurement.
The ratio of two frequencies is arrived at by
using the lower frequency signal for gate control and by having the higher
frequency  signal  counted  as  shown  in  figure  6. 
By  using  the  proper
transducers, ratio measurements may be applied to any phenomena, providing
the phenomena can be represented by sine waves or pulses.  Measurements that
can be performed using the ratio method are clutch slippage, gear ratios and 
frequency divider (or multiplier) operations.
e. Time Interval Measurement.  Time interval measurements are similar
to period measurements; the only exception is that the trigger points on the
single waveform or waveforms are adjustable.  Figure 7 
6