The range of each thermometer.
(1) -38,C to +2,C
(2) -8,C to +32,C
(3) +25,C to +55,C
(4) +50,C to +80,C
(5) +75,C to 105,C
(6) +90,C to 155,C
(7) +145,C to +205,C
(8) +195,C to +305,C
NOTE: There are other sets of thermometers listed in the the supply catalogs, however,
this set is representative of the others.
c. The third type of secondary reference thermometer is the resistance thermometer. The
principle of operation of the resistance thermometer has been discussed before. Here we
have a platinum resistance thermometer and the Mueller Bridge, model E-1003.
6. TEMPERATURE MEASUREMENT SYSTEM TYPE 4100 (Refer to figure 20)
a. INTRODUCTION. The INSTRULAB 4100 has been designed specifically to meet the
electrical, mechanical and environmental specifications of U. S. Army Materiel
Readiness Command, Redstone Arsenal.
b. ELECTRICAL SPECIFICATIONS
(1) Ranges: -75, to 600,F with 0.1,F resolution and -60 to 300,C with 0.1,C
resolution. Ranges shall be selectable with front panel switch.
(2) Display: Minimum four (4) digits plus sign, decimal and unit ID.
(3) Sample Rate: Minimum, 2 readings per second.
(4) Accuracy: ƒ(0.2,F or 0.1,C + 0.1 % of reading) (unit to be matched to
sensor)
(5) Repeatability: